Cowingene HPV (2+16) Detection Kit ( NATBox)

validated specimen: Cervical swab, Urine, Self-collected vaginal Analytes:NATBox;16, 18,(26, 31, 33, 35, 39, 45, 51, 52, 53, 56, 58, 59, 66, 68, 73 and 82)



ORDER INFORMATION

REF

TYPE

PACKAGE

HP01023W

mini II

24tests/kit

INDICATION

in vitro diagnostic medical device

REGULATORY STATUS

CE-IVD

INTENDED USER

For professional use in laboratories with trained staff

TECHNOLOGY

Real-time PCR

TYPE OF ANALYSIS

Qualitative

TARGET SEQUENCE

E6E7,L2L1

ANALYTICAL SPECIFICITY

Human Papillomavirus 16, 18 genotype, Others ( 31, 33, 35, 39, 45, 51, 53, 56, 58, 59, 66, 73, 82)

ANALYTICAL SENSITIVITY  (LoD with probability 95 %)

reaches up to 100 cp/ml with the probability of 95 % (on HPV DNA control, GeneWell)

EXTRACTION/INHIBITION CONTROL

PCR inhibition and DNA extraction efficiency control by Internal Control (IC)

REPORTING UNITS

cp/ml

VALIDATED SPECIMEN

urine,cervical swab

TYPE

cartridge                                

STORAGE

2-8℃

SHELF LIFE

12months

REACTION TIME

80mins

DIAGNOSTIC SPECIFICITY

100.00 % (CI95%: 95.35 % – 100.00 %)

DIAGNOSTIC SENSITIVITY

100.00 % (CI95%: 97.25 % – 100.00 %)

APPLIED INSTRUMENTS

mini II

REQUIRED DETECTION CHANNELS

FAM, HEXROXCY5